86126C UV Light Meter
Specifically designed for 254nm low-pressure mercury lamps
Records UV radiation intensity and dosage
Monitors real-time radiation levels via mobile app
Includes dedicated PC software
The 86126C UV Light Meter is a professional device engineered for measuring the intensity of UVC germicidal lamps. It comes with a 1-meter testing hook to comply with the "WST 367-2012 Regulation of Disinfection Techniques in Healthcare Settings." This ensures accurate measurements based on the 1-meter distance requirement. The device features Bluetooth connectivity, allowing users to monitor lamp intensity remotely through a mobile app—helping to avoid direct UV exposure during testing.




Parameters
Parameter | 86126C UV Light Meter |
---|---|
Spectral Response | 230nm-280nm, λp = 254nm |
Measuring Range | 0 - 20000 μW / cm² |
Resolution | 0.1μW / cm² |
Measuring Accuracy | ±10% |
Sampling Speed | 2048 times/second |
Test Hole Diameter | ¢=10mm |
Power Value Storage Interval | 1s/10s/60s optional |
Recording Period | 10 min / 100 min / 600 min |
Bluetooth Transmission Distance | 50 meters (open space) |
Instrument Size | 148mm * 76mm * 26mm (L*W*H) |
Probe Size | Diameter 39mm * thickness 15mm |
Probe Wire Length | 1 meter |
Weight | About 300 g |
Display | 240*160 dot matrix LCD |
Battery | 4 AAA alkaline dry batteries |


Features
1-Meter Hook Ensures Compliance with Testing Standards
The 86126C is designed to meet the 1-meter testing requirement set forth in the "WST 367-2012 Regulation of Disinfection Techniques in Healthcare Settings" and includes a hook of the appropriate length for accurate positioning.

Wireless Monitoring via Dedicated App
Equipped with Bluetooth functionality, the UVC meter connects to a mobile app for remote, real-time monitoring. This feature helps protect users from harmful UV exposure during testing.

Advanced Data Logging Capabilities
The 86126C offers smart data statistics, including real-time power values, as well as the maximum, minimum, average, and cumulative energy readings. It also logs the measurement time for each session.

Supports Unattended Operation
Users can set an automatic shutdown time between 0–10 minutes. Test data is saved automatically, allowing for unattended measurements. Upon restart, the display shows the most recent data recorded. The device can store up to 25 sets of test records.

the accumulated energy value is automatically counted and showed on the screen when users set the meter to display them

high-precision UVC filters and professional UV sensors reduce measurement errors

Digital signal processing is completed inside the probe, which is not susceptible to interference and provides excellent test accuracy
PC Software for Data Management
The 86126C includes a USB port and specialized PC software. Users can export test data to Excel, generate test reports, and print documentation with ease.

Qualifications and Standards
" WST 367-2012 Regulation of disinfection technique in healthcare settings " A.3 Monitoring of the Effect of Ultraviolet Disinfection
-
UV light meter measurement method
After turning on the UV lamp for 5min, place the UV light meter probe with a measuring wavelength of 253.7nm at the center of 1m vertical distance under the tested UV lamp. The special UV lamp is measured at the recommended distance. After the meter is stable, the displayed value is the irradiance value of the ultraviolet lamp.
-
Judgment of results
For ordinary 30W straight tube UV lamps, the irradiation intensity of the new lamp should meet the requirements of GB 19258; the use of ultraviolet lamp irradiation intensity ≥70μW / cm² is qualified; the 30W high-intensity ultraviolet lamp irradiation intensity ≥180μW / cm² is qualified.
Packing
No. | Item | Quantity | Unit |
---|---|---|---|
1 | 86126C UV Light Meter | 1 | Pcs |
2 | USB Cable | 1 | Pcs |
3 | UV Scale Hooks | 1 | Pcs |
4 | User Manual | 1 | Pcs |
5 | Certificate/Warranty Card | 1 | Pcs |
6 | Plastic Case 34*29*9 cm(L*W*H) | 1 | Pcs |

FAQ
-
1. Can 86126C UV light meter measure UV dose (energy) ?
The 86126C UV light meter can detect the UV radiation intensity and UV dosage (Energy).
-
2. The difference between 86126C and 86125+UVC-X0 UV light meter
1) The radiation intensity measurement range of 86126C is 0-2000μW/cm2, and the radiation intensity measurement range of 86125+UVC-X0 is 0-200000μW/cm2;
2) 86126C is equipped with a 1-meter test hook, which can be connected to the mobile phone APP via Bluetooth and the radiation intensity of the germicidal lamp can be monitored in real time through the APP. 86126C can also be connected to PC software to export test data and print reports. 86125+UVC-X0 UV light meter doesn't have the functions described above.
-
3. Can 86126C UV light meter measure the wavelength of germicidal lamps?
No, the 86126C UV light meter can only measure the UV radiation intensity of the germicidal lamp, and cannot measure the wavelength. If you need to measure the wavelength, you need to use more professional equipment,such as spectrometer.
-
4. Two ways to determine whether it is a 253.7nm germicidal lamp.
1) Although the spectral response range of 86126C or UVC probe is 230-280nm, the measured peak wavelength is 253.7nm, and the response at other wavelengths is very weak. The low-pressure mercury lamp currently known only has a peak wavelength at 253.7nm, if the test data is very low, the germicidal lamp peak wavelength may not be 253.7nm or it's unqualified.
2) Use a spectrometer to measure the spectrum of the germicidal lamp.
-
5. Whether a traceable calibration certificate is included?
The 86126C UV light meter can be ensured to pass the inspection of the Chinese Authoritative Metrology Institute and obtains the calibration certificate (approved by CNAS and ilac-MRA), but the cost of the calibration certificate needs to be paid separately, and the cost is about US $290 (UV intensity and UV dosage inspection).