86125 UV Light Meter +UVALED-X0 Probe

Ideal for UVA LED Light Sources (365nm, 395nm, 405nm, etc.)

Spectral Response: 340nm–420nm (Calibrated at 395nm LED)
Power Measurement Range: 0–200,000 mW/cm²

User Manual

The 86125 digital UV light meter is a versatile device designed to support up to nine different types of probes. Its primary function is to display measurement data. The UVALED-X0 probe, one of the nine compatible probes, features a spectral response range of 340nm to 420nm and is calibrated specifically at 395nm. This makes it highly suitable for measuring the intensity and energy output of UVA LED light sources.

Parameters

Parameters UVALED-X0 Probe
Spectral Response 340nm-420nm, calibrated with 395nm UV LED
Power Measuring Range 0 - 200,000mW / cm²
Resolution 1mW / cm²
Energy Measuring Range 0-9999999mJ/ cm²
Record Time 0-99999s
Measuring Accuracy ±10%
Sampling Speed 2048 times/second
Unit mW / cm²(default), W / cm², W / m²
Test Hole Diameter ¢=10mm
Probe Size 39mm × 32mm × 15mm (L*W*H)
Packing Box Demension 24 * 6 * 5 cm (L * W * H)

Application

Compared to traditional mercury lamps, UVA LED light sources offer a longer lifespan, emit no heat radiation, deliver higher energy output, and provide uniform irradiation. They are safer and more environmentally friendly. These advantages have led to their widespread use in industries such as UV ink curing, UV adhesive curing, PCB exposure, 3D printing, and optical communications. In environments utilizing UVA LED light sources, the 86125 UV light meter paired with the UVALED-X0 probe is essential for accurately measuring radiation intensity and energy.

UV ink curing

UV ink curing

UV adhesive curing

UV adhesive curing

PCB Exposure

PCB Exposure

3D Printing

3D Printing

Features

Broad Spectral Response

The probe is compatible with a wide range of UV LED wavelengths, including 365nm, 375nm, 385nm, 395nm, and 405nm, for precise measurement of light intensity and energy.

Simultaneous Measurement of Energy and Power

The UV radiometer can simultaneously record both energy and power values. It also includes comprehensive statistical functions, offering real-time power readings along with maximum, minimum, and average values, measurement duration, and total energy value.

High-Speed Data Acquisition

With a sampling rate of up to 2048 data points per second, the meter ensures highly accurate and reliable measurement results.

Easily Switchable Measurement Units

The device supports three interchangeable units: mW/cm², W/cm², and W/m². Unit conversions are automatic, eliminating the need for manual calculations.

Advanced Digital Probe Technology

Leveraging cutting-edge digital probe technology, all data collection and processing are performed directly within the probe. Digital signal transmission enhances resistance to interference, ensuring more stable and consistent measurement data.

FAQ

  1. 1. The difference between UVALED-X0, UVALED-X1 and UVALED-X3 UV probe for 86125 UV light meter host

    1) The test aperture of the UVALED-X0 probe is 10mm, which is mainly used to detect UVA LED plane light sources in curing and exposure industries;

    2) The test aperture of the UVALED-X1 probe is 1mm, which is mainly used to detect UVA LED point light sources in curing and exposure industries;

    3) The UVALED-X3 probe is mainly used for the detection of low-power UVA LED light sources, such as nail art lamps.

  2. 2. What's the working temperature of the UV probes?

    The 9 probes of the 86125 UV light meter are not heat-resistant. The probe can be used for a long time at 80℃ or lower. If the temperature is higher than 80℃, it is recommended to use UV energy meter for UVA light source detection.

  3. 3. Which probe is recommended if I need to measure both mercury lamp and LED light source?

    It is recommended to use the 86125 UV light meter host and UVA-X0, UVALED-X0 probes.

  4. 4. Whether a traceable calibration certificate is included?

    UVA-X0 and UVA-X1 are mainly used to detect 365nm high-pressure mercury lamps, which can be ensured to pass the inspection of the Metrology Institute and obtain a calibration certificate (approved by CNAS and ilac-MRA). The cost of the calibration certificate is about US$200 ; but the UVALED-X0, UVALED-X1 and UVALED-X3 probes are professionally used to detect UVA LED light sources. There is no unified inspection standard for LED light sources, each light source and UV energy meter manufacturer has their own calibration standards. We can provide the factory calibration reports for these three probes special for UV LED, but no calibration certificates can be provided.